Z-2627-2016 Class II Terminated
FDA device recall Z-2627-2016 was initiated by Merge Healthcare, Inc. on Jan 31, 2016 and is designated Class II. Reason for recall: Potential for RadSuite AV viewer to skip image slices when the user presses page up or page down when the viewer displays multiple viewports and the user has mouse focused on any viewport other than… The recall status is terminated (terminated Feb 10, 2017). Affected quantity: 25 sites potentially have the affected versions.
Recall details
- Product Type
- Devices
- Report Date
- Aug 31, 2016
- Initiation Date
- Jan 31, 2016
- Termination Date
- Feb 10, 2017
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 25 sites potentially have the affected versions
Product description
RadSuite software. The firm name on the label is Merge Healthcare, Hartland, WI.
Reason for recall
Potential for RadSuite AV viewer to skip image slices when the user presses page up or page down when the viewer displays multiple viewports and the user has mouse focused on any viewport other than the first viewport.
Distribution pattern
Distribution was made to AL, AZ, CA, IL, MD, MA, MI, MO, NV, NJ, NC, OH, PA, TN, and VA. Government distribution was made to OK. There was no foreign/military distribution.
Code information
Versions: 5.30.7, 5.30.7 HF0429E, 5.30.7 HF0441E, 5.30.7 HF0446E, 5.30.7 HF0470E, 5.30.7 HF0473E, 5.30.8, 5.30.8 HF0477E, 5.30.8 HF0490E, 5.30.8 HF0495 SB0508E, 5.30.8 HF0495E, 5.30.8 HF0495E, SB504E, 5.30.8 HF0550, 5.30.8 SB0485E, 5.30.8 SB0486E, 5.30.X SB0381E, 5.35.1, 5.35.1 HF0487E, 5.35.1 HF0489E, 5.35.2, 5.35.3, 5.35.4, and 5.35.4 HF0513