FDA device recall Z-2628-2017 was initiated by Merge Healthcare, Inc. on Apr 15, 2016 and is designated Class II. Reason for recall: There are potential issues with results reporting for certain run-based tests. Under certain conditions, the wrong results could inadvertently be verified. The recall status is terminated (terminated Mar 13, 2019). Affected quantity: 638 sites have the potentially affected versions.
Recall details
- Product Type
- Devices
- Report Date
- Jul 5, 2017
- Initiation Date
- Apr 15, 2016
- Termination Date
- Mar 13, 2019
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 638 sites have the potentially affected versions
Product description
Merge LIS software. The firm name on the label is Merge Healthcare. Merge LIS system is a complete system for ordering, managing and reporting a patient s laboratory work, from the time of order entry to the time the laboratory test results are reported.
Reason for recall
There are potential issues with results reporting for certain run-based tests. Under certain conditions, the wrong results could inadvertently be verified.
Distribution pattern
Nationwide Distribution to facilities and distributors. Distribution was made to the Virgin Islands. Foreign distribution was also made. There was no military/government distribution.
Code information
Versions V3.7, 3.7.1, 3.7.2, 3.7.3, 3.8, 3.8.0.1, 3.8.1, 3.8.2, 3.9.1, 3.9.2, 3.9.3, 4.1, 4.1.1, 4.1.2, 4.1.3, and 4.1.4