Z-2660-2023 Class II Ongoing

Recalled by Olympus Corporation of the Americas — Center Valley, PA

Recall Details

Product Type
Devices
Report Date
October 11, 2023
Initiation Date
August 10, 2023
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
6484 units

Product Description

Veran Endobronchial: Models: INS-5925 SPiN Access Catheter¿ 180 Degree for use w/ Pentax Scopes INS-5920 SPiN Access Catheter¿ 90 Degree for use w/ Pentax Scopes INS-5915 SPiN Access Catheter¿ 180 for use w/ Olympus Scopes INS-5910 SPiN Access Catheter¿ 90 for use w/ Olympus 190 Scopes INS-5905 SPiN Access Catheter¿ 180 for use w/ Olympus Scopes INS-5900 SPiN Access Catheter¿ 90 Degree for use w/ Olympus Scopes INS-5700 SPiN EWC Biopsy Guide Kit

Reason for Recall

Non-conformances in Veran devices (ig4 Image Guided System, SPiN Thoracic Navigation System, SPiN Planning Laptop Workstation, and all associated navigation instruments) manufacturing to design requirements and insufficient evidence to demonstrate adequate electromagnetic compatibility (EMC). EMC is the ability of electrical equipment and systems to function acceptably without interference and to not interfere with other devices within its operating environment

Distribution Pattern

Worldwide distribution - US Nationwide and the countries of CN, DE, HK, IT, KR, SG, TH, TW.

Code Information

Model- UDI: INS-5925 SPiN Access Catheter¿ 180 Degree for use w/ Pentax Scopes 00815686021313; INS-5920 SPiN Access Catheter¿ 90 Degree for use w/ Pentax Scopes 00815686021306; INS-5915 SPiN Access Catheter¿ 180 for use w/ Olympus Scopes 00815686021290; INS-5910 SPiN Access Catheter¿ 90 for use w/ Olympus 190 Scopes 00815686021276; INS-5905 SPiN Access Catheter¿ 180 for use w/ Olympus Scopes 00815686021269; INS-5900 SPiN Access Catheter¿ 90 Degree for use w/ Olympus Scopes 00815686021153; INS-5700 SPiN EWC Biopsy Guide Kit 00815686021108. All serial numbers and all lot numbers