Z-2691-2016 Class II Terminated

Recalled by Elekta, Inc. — Atlanta, GA

FDA device recall Z-2691-2016 was initiated by Elekta, Inc. on Aug 18, 2016 and is designated Class II. Reason for recall: Potentially unrecognized incorrect position of the treatment couch in 3D workflow, i.e. the HexaPOD has not moved fully to the 3D position. The recall status is terminated (terminated May 23, 2017). Affected quantity: 13.

Recall details

Product Type
Devices
Report Date
Sep 7, 2016
Initiation Date
Aug 18, 2016
Termination Date
May 23, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
13

Product description

HexaPOD evo RT System Product Usage: The intended use of the device is the control of accurate patient positioning with assistance of a 3D Tracking System in a radiotherapy environment.

Reason for recall

Potentially unrecognized incorrect position of the treatment couch in 3D workflow, i.e. the HexaPOD has not moved fully to the 3D position.

Distribution pattern

Worldwide Distribution - US Nationwide in the states of LA, PA, WA, and the countries of: Australia, Denmark, France, India, Italy, and Japan.

Code information

iGuide 2.2.0