Z-2697-2017 Class II Terminated

Recalled by Elekta, Inc. — Atlanta, GA

FDA device recall Z-2697-2017 was initiated by Elekta, Inc. on May 23, 2017 and is designated Class II. Reason for recall: If the gantry module is replaced after original installation the configuration settings may be missing. The identified risk for this issue is electrical safety for technicians doing maintenance on the… The recall status is terminated (terminated Oct 18, 2017). Affected quantity: 2.

Recall details

Product Type
Devices
Report Date
Jul 19, 2017
Initiation Date
May 23, 2017
Termination Date
Oct 18, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
2

Product description

Leksell Gamma Knife Icon is a teletherapy device intended for stereotactic irradiation of head structures ranging from very small target sizes of a few millimeters to several centimeters e.g., metastatic tumors, recurrent glioblastomas, trigeminal neuralgia, medically refractory essential tremor, orbital tumors, ocular tumors, optic nerve tumors, benign diseases (such as meningiomas, vestibular schwannomas, post-surgical pituitary adenomas, craniopharyngioma, hemangioblastomas, schwannomas, arteriovenous malformations, cavernous malformations, chordomas, glomus tumors, hemangiomas), skull base tumors, head and neck tumors (such as unknown primary of the head and neck, oral cavity, hypopharynx, oropharynx, nasopharynx, sinonasal, salivary gland), and pediatric tumors (such as glioma, ependymoma, pituitary tumors, hemangioblastoma, craniopharyngioma, meningioma, metastasis, medulloblastoma, nasopharyngeal tumors, arteriovenous malformations, cavernous malformations, skull base tumors).

Reason for recall

If the gantry module is replaced after original installation the configuration settings may be missing. The identified risk for this issue is electrical safety for technicians doing maintenance on the X-ray generator.

Distribution pattern

US Distribution to the states of : NY and PA.

Code information

Serial Numbers 6018, 6047, 6090, 6094, 6171, 8018

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2697-2017
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-2697-2017. Report a data issue: hello@regdatalab.com.