Z-2715-2016 Class II Terminated

Recalled by Merge Healthcare, Inc. — Hartland, WI

FDA device recall Z-2715-2016 was initiated by Merge Healthcare, Inc. on Apr 9, 2015 and is designated Class II. Reason for recall: When RadSuite is used with IPID (Issuer of Patient ID) as a part of the "Patient Identifier," it is possible in some circumstances that the demographics of one patient will be applied to a study or… The recall status is terminated (terminated Aug 24, 2017). Affected quantity: 10 sites have the affected software that is configured with the specific conditions listed in Code Information.

Recall details

Product Type
Devices
Report Date
Sep 14, 2016
Initiation Date
Apr 9, 2015
Termination Date
Aug 24, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
10 sites have the affected software that is configured with the specific conditions listed in Code Information

Product description

Merge RadSuite software. Radiological image processing system.

Reason for recall

When RadSuite is used with IPID (Issuer of Patient ID) as a part of the "Patient Identifier," it is possible in some circumstances that the demographics of one patient will be applied to a study or studies for another patient.

Distribution pattern

Distribution was made to medical facilities in AL, MI, MO, PA, TN, and TX.

Code information

RadSuite Versions: 5.30.8, 5.35.3, 5.35.4, 5.35.5, 8.30.7.3, 8.30.7.4, 8.30.7.5, 8.30.7.7 are affected, BUT ONLY if they are configured with all of the following conditions: (1) Must have IPID enabled; (2) Must use IPID morphers (3) Must Define a Defaultvalue in the IPID Morpher; (4) Must use the Aggregating Morpher; (5) Must not set OverwritelfPresent=false; and (6) Must be on an internal store (this is data that is already present on EA being stored to Radsuite, this excludes direct DICOM stores to the EA).

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2715-2016
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-2715-2016. Report a data issue: hello@regdatalab.com.