Z-2746-2017 Class II Terminated

Recalled by Medtronic Navigation, Inc. — Louisville, CO

FDA device recall Z-2746-2017 was initiated by Medtronic Navigation, Inc. on May 10, 2017 and is designated Class II. Reason for recall: Software issue related to the StealthStation S7 system and the Synergy Spine application Version 2.1 configured with Spine Tool Install CD version 25. Issue may result in user being unable to… The recall status is terminated (terminated Jan 8, 2021). Affected quantity: 108.

Recall details

Product Type
Devices
Report Date
Jul 26, 2017
Initiation Date
May 10, 2017
Termination Date
Jan 8, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
108

Product description

Install CD, Spine tools, Plus and S7, Version 25. Model Number 9731958.

Reason for recall

Software issue related to the StealthStation S7 system and the Synergy Spine application Version 2.1 configured with Spine Tool Install CD version 25. Issue may result in user being unable to navigate the Navigated Elevate Inserter with the StealthAiR Spine Frame during spine surgical procedures.

Distribution pattern

Worldwide Distribution - USA (nationwide) Distribution and to the countries of : Brazil, Germany, Switzerland and Great Britain.

Code information

Kits that include affected 9731958 INSTALL CD: INST SET 9734833 UNIVERSAL NAVLOCK, INST SET 9734834 CAPSTONE TRIALS, INST SET 9734835 CLYDESDALE TRIALS, INST SET 9734836 TLIF/DLIF INSERTER, INST SET 9735283 SOLERA 5.5/6.0 DRIVERS, INST SET 9735284 SOLERA NONCAN TAPS, INST SET 9735285 SOLERA ILIAC TAPS, TAPS KIT 9735465 SOLERA AWL-TIP TAPS, REF SET 9735500 SPINE REFRENCING, UPGRD KIT 9735522 SPINE REVISION, INSTR SET 9735607 VERTEX SELECT

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2746-2017
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Medtronic Navigation, Inc.? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Medtronic Navigation, Inc.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Medtronic Navigation, Inc. →

Source: openFDA record for Z-2746-2017. Report a data issue: hello@regdatalab.com.