Z-2795-2018 Class II Terminated

Recalled by Hamilton Medical AG — Bonaduz

FDA device recall Z-2795-2018 was initiated by Hamilton Medical AG on Mar 9, 2018 and is designated Class II. Reason for recall: An out of specification component on the control board may cause the ventilator to open the inspiratory channel and the expiratory valve. The ventilator will shut down and an emergency buzzer… The recall status is terminated (terminated Mar 18, 2019). Affected quantity: 16 devices.

Recall Details

Product Type
Devices
Report Date
Aug 22, 2018
Initiation Date
Mar 9, 2018
Termination Date
Mar 18, 2019
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
16 devices

Product Description

Hamilton-C1 Ventilator Unit Product Usage: The HAMILTON-C1 ventilator is intended to provide positive pressure ventilation support to adults and pediatrics. The device is used in the in the intensive care ward, in the recovery room, and during transfer of ventilated patients within the hospital.

Reason for Recall

An out of specification component on the control board may cause the ventilator to open the inspiratory channel and the expiratory valve. The ventilator will shut down and an emergency buzzer providing an acoustic alarm and the blinking alarm LED on the device s front side.

Distribution Pattern

US distribution to CA, CO, IL, LA, MN, NY, OK, VA, WI. Worldwide distribution to China, Germany, India, Indonesia, Israel, Kazakhstan, Korea, Pakistan, Taiwan, Poland, Armenia.

Code Information

Serial Number: 12244, 12246, 12338, 12343, 12344, 12345, 12351, 12352, 12355, 12370, 12380, 12382, 12383, 12412, 12426, 12432

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2795-2018
openFDA file date
Oct 26, 2026
Loaded by RegDataLab
Oct 27, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.