Z-2528-2023 Class I Ongoing

Recalled by Hamilton Medical, Inc. — Reno, NV

FDA device recall Z-2528-2023 was initiated by Hamilton Medical, Inc. on Jul 26, 2023 and is designated Class I. Reason for recall: Degrading capacitors on the control board of ventilators and spare parts may leak electrolyte onto the control board causing a short circuit on the board and/or the capacitor to lose function and lead… The recall status is ongoing. Affected quantity: 12 units.

Recall details

Product Type
Devices
Report Date
Sep 27, 2023
Initiation Date
Jul 26, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
12 units

Product description

Hamilton-MR1, REF: 161010, Intensive Care Ventilator

Reason for recall

Degrading capacitors on the control board of ventilators and spare parts may leak electrolyte onto the control board causing a short circuit on the board and/or the capacitor to lose function and lead to interruption of ventilation.

Distribution pattern

Distribution US nationwide, including Puerto Rico. OUS: Not provided

Code information

UDI: 07630002800761 & 07630002813556 / Affected Serial Numbers: 3422 3423 3424 3425 3426 3427 3428 3429 3469 3503 3510 3800

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2528-2023
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Hamilton Medical, Inc.? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Hamilton Medical, Inc.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Hamilton Medical, Inc. →

Source: openFDA record for Z-2528-2023. Report a data issue: hello@regdatalab.com.