Z-2167-2025 Class I Ongoing

Recalled by Hamilton Medical AG — Domat/Ems

FDA device recall Z-2167-2025 was initiated by Hamilton Medical AG on Jun 26, 2025 and is designated Class I. Reason for recall: Ventilator coaxial breathing circuit may have a crack in blue tubing, which is used to deliver fresh gas to the patient. The crack will compromise inspiratory and expiratory flows, leading to partial… The recall status is ongoing. Affected quantity: 2,560.

Recall details

Product Type
Devices
Report Date
Aug 6, 2025
Initiation Date
Jun 26, 2025
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
2,560

Product description

Breathing circuit set, coaxial, 1.80 m, with flow sensor and expiratory valve set, REF: 260128, used with HAMILTON-C1, HAMILTON-T1 and HAMILTON-MR1 ventilators.

Reason for recall

Ventilator coaxial breathing circuit may have a crack in blue tubing, which is used to deliver fresh gas to the patient. The crack will compromise inspiratory and expiratory flows, leading to partial or complete rebreathing of exhaled gases, which could lead to acute hypercapnia, respiratory acidosis, and if unrecognized organ dysfunction.

Distribution pattern

US Nationwide distribution in the states of NH, NY, CT, PA, IL, ME, CO, NC, NJ, UT, TX, GA, SC, VA, IN, SD, MO, OR, TN, MA, AZ, NM.

Code information

UDI-DI: 07630002802963, Lots: 199675, 199676

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2167-2025
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Hamilton Medical AG? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Hamilton Medical AG: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Hamilton Medical AG →

Source: openFDA record for Z-2167-2025. Report a data issue: hello@regdatalab.com.