Z-2020-2024 Class I Ongoing
FDA device recall Z-2020-2024 was initiated by Hamilton Medical AG on May 15, 2024 and is designated Class I. Reason for recall: Ventilator may enter sensor fail mode, ventilation may not re-initiate, after patient is reconnected, which may cause hypoxia, if this sequence occurs: 1)User presses O2 enrichment, 2)User disconnects… The recall status is ongoing. Affected quantity: 68.
Recall details
- Product Type
- Devices
- Report Date
- Jun 26, 2024
- Initiation Date
- May 15, 2024
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 68
Product description
Reason for recall
Distribution pattern
US Nationwide distribution in the states of GA, KY, FL, IN, NV, MI, VA, OR, TX, SC, CA, PA.