Z-2020-2024 Class I Ongoing

Recalled by Hamilton Medical AG — Bonaduz, N/A

Recall Details

Product Type
Devices
Report Date
June 26, 2024
Initiation Date
May 15, 2024
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
68

Product Description

Ventilator HAMILTON-C6, PN: 160021

Reason for Recall

Ventilator may enter sensor fail mode, ventilation may not re-initiate, after patient is reconnected, which may cause hypoxia, if this sequence occurs: 1)User presses O2 enrichment, 2)User disconnects endotracheal tube for open suctioning, 3)Sensor error initiated, sensor fail mode occurs, 4)Patient is re-connected with sensor fail mode active, 5) Ventilation is not re-initiated by the ventilator.

Distribution Pattern

US Nationwide distribution in the states of GA, KY, FL, IN, NV, MI, VA, OR, TX, SC, CA, PA.

Code Information

UDI: 07630002808590, Software Versions: SW v1.1.4, SW v1.1.5, SW v1.1.6