Z-2796-2018 Class II Terminated

Recalled by Hamilton Medical AG — Bonaduz

Recall Details

Product Type
Devices
Report Date
August 22, 2018
Initiation Date
March 9, 2018
Termination Date
March 18, 2019
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
2 devices

Product Description

Hamilton-C3 Ventilator Unit Product Usage: The HAMILTON-C3 ventilator is intended to provide positive pressure ventilator support to adults, pediatrics, infants and neonates depending on its configuration. The device is used in the intensive care ward, intermediate care ward, emergency ward, long term acute care hospital or in the recovery room, and during transfer of ventilated patients within the hospital.

Reason for Recall

An out of specification component on the control board may cause the ventilator to open the inspiratory channel and the expiratory valve. The ventilator will shut down and an emergency buzzer providing an acoustic alarm and the blinking alarm LED on the device s front side.

Distribution Pattern

US distribution to CA, CO, IL, LA, MN, NY, OK, VA, WI. Worldwide distribution to China, Germany, India, Indonesia, Israel, Kazakhstan, Korea, Pakistan, Taiwan, Poland, Armenia.

Code Information

Serial Number: 3084, 3088