Z-2858-2020 Class II Terminated

Recalled by Globus Medical, Inc. — Audubon, PA

FDA device recall Z-2858-2020 was initiated by Globus Medical, Inc. on Jul 22, 2020 and is designated Class II. Reason for recall: Specific lots of ALTERA¿ implants have internal components that may have been manufactured using stainless steel rather than the specified cobalt chromium molybdenum alloy. The recall status is terminated (terminated Aug 19, 2021).

Recall details

Product Type
Devices
Report Date
Sep 2, 2020
Initiation Date
Jul 22, 2020
Termination Date
Aug 19, 2021
Voluntary/Mandated
Voluntary: Firm initiated

Product description

ALTERA Spacer ALTERA Spacer, 10x31, 8-12mm, 8¿

Reason for recall

Specific lots of ALTERA¿ implants have internal components that may have been manufactured using stainless steel rather than the specified cobalt chromium molybdenum alloy.

Distribution pattern

Worldwide Distribution

Code information

Part #: 1124.1111; Lot # / UDI #: BAX156ME / (01)00889095069679(10)BAX156ME; BAX156NE / (01)00889095069679(10)BAX156NE; BAX167PE / (01)00889095069679(10)BAX167PE; BAX229VE / (01)00889095069679(10)BAX229VE; BAX290TE / (01)00889095069679(10)BAX290TE; BAX290UE / (01)00889095069679(10)BAX290UE; BAX294WE / (01)00889095069679(10)BAX294WE; BAX376DE / (01)00889095069679(10)BAX376DE; BAX376CE / (01)00889095069679(10)BAX376CE

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2858-2020
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Globus Medical, Inc.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-2858-2020. Report a data issue: hello@regdatalab.com.