Z-2916-2026 Class II Ongoing

Recalled by Spacelabs Healthcare, Inc. — Snoqualmie, WA

FDA device recall Z-2916-2026 was initiated by Spacelabs Healthcare, Inc. on June 4, 2026 and is designated Class II. Reason for recall: Due to off-the-shelf display port component that has audio/video failures The recall status is ongoing. Affected quantity: 6074.

Recall Details

Product Type
Devices
Report Date
August 19, 2026
Initiation Date
June 4, 2026
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
6074

Product Description

Brand Name: Xhibit Product Name: Xhibit Central Station Model/Catalog Number: 96102 Software Version: N/A Product Description: The Spacelabs Healthcare (Spacelabs) Xhibit Central Station, Model 96102. Xhibit offers remote surveillance of patient data for those patients connected to a Spacelabs Healthcare bedside monitors and telemetry systems. The DisplayPort audio/video splitter allows users to extend the audio/video source from the Xhibit to additional external displays or monitors. This extension of audio/video is for viewing only and does not allow interaction with the Xhibit. Component: The DisplayPort audio/video splitter is a 3rd party external accessory to Xhibit.

Reason for Recall

Due to off-the-shelf display port component that has audio/video failures

Distribution Pattern

Worldwide - US Nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, VA, WA, WI, WY and the countries of CANADA, FRANCE, ITALY, MALAYSIA, THAILAND.

Code Information

Model/Catalog Number: 96102 DisplayPort Splitter Part Number: 010-1975-00 UDI-DI codes: 1084152212706 10841522100345 Serial Numbers: Pending