Z-2959-2026 Class II Ongoing

Recalled by Tobii Dynavox Llc — Pittsburgh, PA

FDA device recall Z-2959-2026 was initiated by Tobii Dynavox Llc on July 8, 2026 and is designated Class II. Reason for recall: Under certain conditions, the battery mounting tabs may become damaged, allowing the battery to move within the device enclosure, potentially leading to an internal short circuit. This may cause local… The recall status is ongoing.

Recall Details

Product Type
Devices
Report Date
August 26, 2026
Initiation Date
July 8, 2026
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity

Product Description

TD Navio Midi Speech Generating Device Model/Catalog Number: 1000798

Reason for Recall

Under certain conditions, the battery mounting tabs may become damaged, allowing the battery to move within the device enclosure, potentially leading to an internal short circuit. This may cause localized heating and gas venting within the device enclosure.

Distribution Pattern

US Nationwide distribution.

Code Information

UDI-DI 073400747602056 Serial Numbers TDNM1-240628000004 through TDNM1-260617070001. Corrected/future production is differentiated by new serialization from TDNM1-260617070001 onward (production batch 20260615, improved design).