Z-2960-2026 Class II Ongoing

Recalled by Tobii Dynavox Llc — Pittsburgh, PA

FDA device recall Z-2960-2026 was initiated by Tobii Dynavox Llc on Jul 8, 2026 and is designated Class II. Reason for recall: Under certain conditions, the battery mounting tabs may become damaged, allowing the battery to move within the device enclosure, potentially leading to an internal short circuit. This may cause… The recall status is ongoing.

Recall details

Product Type
Devices
Report Date
Aug 26, 2026
Initiation Date
Jul 8, 2026
Voluntary/Mandated
Voluntary: Firm initiated

Product description

TD Navio Maxi Speech Generating Device Model/Catalog Number: 1000799

Reason for recall

Under certain conditions, the battery mounting tabs may become damaged, allowing the battery to move within the device enclosure, potentially leading to an internal short circuit. This may cause localized heating and gas venting within the device enclosure.

Distribution pattern

US Nationwide distribution.

Code information

UDI-DI 073400747602070 Serial Numbers TDNL1-240904000016 through TDNL1-260617070001. Corrected/future production is differentiated by new serialization from TDNL1-260617070001 onward (production batch 20260615, improved design).

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2960-2026
openFDA file date
Nov 24, 2026
Loaded by RegDataLab
Nov 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Tobii Dynavox Llc? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-2960-2026. Report a data issue: hello@regdatalab.com.