Z-3056-2017 Class II Terminated
Recalled by Medtronic Navigation, Inc.
— Louisville, CO
FDA device recall Z-3056-2017 was initiated by Medtronic Navigation, Inc. on Jun 29, 2017 and is designated Class II. Reason for recall: The instruments are not able to pass instrument verification which is performed prior to the use of each instrument. The recall status is terminated (terminated Aug 3, 2018). Affected quantity: 15 kits.
Recall details
- Product Type
- Devices
- Report Date
- Aug 30, 2017
- Initiation Date
- Jun 29, 2017
- Termination Date
- Aug 3, 2018
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 15 kits
Product description
Medtronic Instrument Kit 9733908 Fusion ENT, reusable, RX. This kit includes Kit 9733452 (containing 9733449, Straight Suction, 9733450, 70 deg Curved Suction, and 9733451, 90 deg Curved Suction). The firm name on the label is Medtronic Navigation, Inc., Louisville, CO. The suction devices are part of a family of reusable ENT instruments that are used with the StealthStationTM AxiEM surgical navigation system. The AxiEM system is an electromagnetic (EM) tracking solution for surgical tools. The system measures an induced voltage created by local magnetic fields from a set of coils in the transmit coil array (TCA). Each surgical tool has a set of receiving coils to sense the magnetic field strength. Using a mathematical algorithm, the position and orientation of the instrument can be computed and overlaid/registered to imagery on a computer monitor, resulting in the ability to track the instrument tip in relation to the patients anatomy. AxiEM ENT Suction devices are used as suction and pointing devices. The suction devices are intended to have an ENT Instrument Tracker mounted on it
Reason for recall
The instruments are not able to pass instrument verification which is performed prior to the use of each instrument.
Distribution pattern
Worldwide Distribution - US (nationwide) and Internationally to Canada.
Code information
Each of the devices within the kit are labeled with information specific to each of the reusable devices within the kit and those devices are what goes into the customer inventory. The outer box where the lot number is printed is discarded by the customer on receipt and the separate devices within the box are what is stored on the shelf. Therefore, no kit lot number is being provided.
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-3056-2017
- openFDA file date
- Nov 25, 2026
- Loaded by RegDataLab
- Nov 26, 2026
These dates say when openFDA published the data file and when RegDataLab last loaded it.
Need the full FDA record for Medtronic Navigation, Inc.? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.
Question to ask in Claude
Use RegDataLab to give me an FDA risk summary for Medtronic Navigation, Inc.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.
Ask Claude about Medtronic Navigation, Inc. →
Use it in ChatGPT
Get a free API key
How the connector works