Z-3075-2017 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- September 6, 2017
- Initiation Date
- July 12, 2017
- Termination Date
- December 16, 2020
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 1866 units
Product Description
VITEK¿ 2 Compact 15, REF numbers: 27415 and 27415R software and reagent cards designed for the identification (ID) and antimicrobial susceptibility testing (AST) of bacteria and yeast.
Reason for Recall
Customers have reported that some VITEK¿ 2 cards are staying in preliminary status, not finalizing after ejection from the instrument, and not allowing cards in subsequent carousel slots to be processed. The issue was reported to occur on VITEK¿ 2 Compact 15 and Compact 30 systems following a system software update to version 8.01.
Distribution Pattern
Worldwide
Code Information