Z-3127-2017 Class II Terminated

Recalled by Orthofix, Inc — Lewisville, TX

FDA device recall Z-3127-2017 was initiated by Orthofix, Inc on Aug 2, 2017 and is designated Class II. Reason for recall: The plastic core of the Decorticating Planer was found to be missing, which could result in scoring of the bone screw neck if used at or beyond a 23 degree angle. The recall status is terminated (terminated May 4, 2018). Affected quantity: 6 planers.

Recall details

Product Type
Devices
Report Date
Sep 20, 2017
Initiation Date
Aug 2, 2017
Termination Date
May 4, 2018
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
6 planers

Product description

Firebird NXG Spinal Fixation System Decorticating Planer, Catalog #36-1334, non-sterile, Rx. .A Decorticating Planer is used to decorticate bone surrounding the spherical head of an implanted bone screw in order to provide clearance to allow for proper seating of the modular body, as well as to provide full range of modular body angulation. It is placed axially over the spherical head of the bone screw and rotated clockwise and counterclockwise to achieve the needed decortication

Reason for recall

The plastic core of the Decorticating Planer was found to be missing, which could result in scoring of the bone screw neck if used at or beyond a 23 degree angle.

Distribution pattern

Distribution was made to distributors located in CA, FL, and RI. There was no foreign/military/government distribution.

Code information

Lot #203013-TD26

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-3127-2017
openFDA file date
Nov 24, 2026
Loaded by RegDataLab
Nov 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Orthofix, Inc? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-3127-2017. Report a data issue: hello@regdatalab.com.