Z-3203-2026 Class II Ongoing

Recalled by Howmedica Osteonics Corp. — Mahwah, NJ

FDA device recall Z-3203-2026 was initiated by Howmedica Osteonics Corp. on Jul 30, 2026 and is designated Class II. Reason for recall: Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which… The recall status is ongoing. Affected quantity: 12,156 US.

Recall details

Product Type
Devices
Report Date
Sep 23, 2026
Initiation Date
Jul 30, 2026
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
12,156 US

Product description

SCORPIO RECESSED METAL BACKED PATELLA 3045-0026 3045-0028 3045-0030 3045-0032 Orthopaedic implant components manufactured from UHMWPE (polyethylene)

Reason for recall

Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.

Distribution pattern

Worldwide distribution - US Nationwide and the countries of Canada, LATAM, Europe, Asia Pacific, and South Africa.

Code information

3045-0026 No UDI Lots 12WAHA 12WAHB HOHE HOVX 3045-0028 No UDI Lots 13VAW 13VAWA 3045-0030 No UDI Lots BIYIE HNZY 3045-0032 No UDI Lots BIYIH HOYK

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-3203-2026
openFDA file date
Nov 24, 2026
Loaded by RegDataLab
Nov 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Howmedica Osteonics Corp.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-3203-2026. Report a data issue: hello@regdatalab.com.