Z-0048-2014 Class II Terminated

Recalled by Abbott Medical Optics, Inc. — Milpitas, CA

FDA device recall Z-0048-2014 was initiated by Abbott Medical Optics, Inc. on Sep 16, 2013 and is designated Class II. Reason for recall: This corrective action will address an identified failure of the lamp and illumination system within the laser unit caused by an unexpected failure of a thyristor component on the AC Power… The recall status is terminated (terminated Nov 12, 2014). Affected quantity: 834 units in total.

Recall details

Product Type
Devices
Report Date
Oct 30, 2013
Initiation Date
Sep 16, 2013
Termination Date
Nov 12, 2014
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
834 units in total

Product description

STAR Excimer Laser System with Variable Spot Scanning, Model: STAR S4, Catalog Number: 0030-4077, Product is manufactured and distributed by AMO Manufacturing USA, LLC, Milpitas, CA The STAR Excimer Laser Systems are indicated for laser assisted in situ keratomileusis (LASIK), photorefractive keratectomy (PRK), and phototherapeutic keratectomy (PTK) eye surgery in certain patient populations.

Reason for recall

This corrective action will address an identified failure of the lamp and illumination system within the laser unit caused by an unexpected failure of a thyristor component on the AC Power Distribution Printed Circuit Board (PCB).

Distribution pattern

Worldwide distribution: US (nationwide) including Puerto Rico and countries of: Asia Pacific, Canada, Europe, Africa, Middle East, Japan, and Latin America.

Code information

All units of these models: STAR S2 Excimer Laser System; STAR S3 ActiveTrak; STAR S4 with Variable Spot Scanning; and STAR S41R; There are no expiration dates for the STAR Excimer Laser Systems.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0048-2014
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Abbott Medical Optics, Inc.? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Abbott Medical Optics, Inc.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Abbott Medical Optics, Inc. →

Source: openFDA record for Z-0048-2014. Report a data issue: hello@regdatalab.com.