Z-0049-2014 Class II Terminated
FDA device recall Z-0049-2014 was initiated by Abbott Medical Optics, Inc. on Sep 16, 2013 and is designated Class II. Reason for recall: This corrective action will address an identified failure of the lamp and illumination system within the laser unit caused by an unexpected failure of a thyristor component on the AC Power… The recall status is terminated (terminated Nov 12, 2014). Affected quantity: 834 units in total.
Recall details
- Product Type
- Devices
- Report Date
- Oct 30, 2013
- Initiation Date
- Sep 16, 2013
- Termination Date
- Nov 12, 2014
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 834 units in total
Product description
Reason for recall
Distribution pattern
Worldwide distribution: US (nationwide) including Puerto Rico and countries of: Asia Pacific, Canada, Europe, Africa, Middle East, Japan, and Latin America.