Z-1670-2013 Class II Terminated

Recalled by Abbott Medical Optics, Inc. — Milpitas, CA

FDA device recall Z-1670-2013 was initiated by Abbott Medical Optics, Inc. on May 23, 2013 and is designated Class II. Reason for recall: Abbott Medical Optics Inc. (AMO) initiated this Urgent Notice of Medical Device Correction and Advisory Notice to all customers who use the AMO Sovereign¿ Compact Phacoemulsification System, Version… The recall status is terminated (terminated Oct 6, 2014). Affected quantity: 534 systems.

Recall details

Product Type
Devices
Report Date
Jul 17, 2013
Initiation Date
May 23, 2013
Termination Date
Oct 6, 2014
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
534 systems

Product description

Sovereign¿ Compact Phacoemulsification System, v 5.3, Part Number: CMP680300. The system is used for cataract surgery.

Reason for recall

Abbott Medical Optics Inc. (AMO) initiated this Urgent Notice of Medical Device Correction and Advisory Notice to all customers who use the AMO Sovereign¿ Compact Phacoemulsification System, Version 5.3, because it has the potential to lose the vacuum tone mode during surgery.

Distribution pattern

Worldwide Distribution, including Nationwide (US) and countries in Latin America, EMEA, and APAC.

Code information

Product Name: Sovereign¿ Compact Console Universal, Part Number: CMP680300. The Sovereign¿ Compact console is identified with a unique, sequentially assigned 3 digit number identified below as XXX. For Sovereign¿ Compact systems manufactured at the Third Party Manufacturer, Sanmina-SCI Corporation, the serial number format is YYYY35XXX where Y is for the year manufactured and "35" designates manufacturing at Sanmina-SCI Corporation. The Sanmina-SCI Corporation started manufacturing the Sovereign¿ Compact v 5.3 systems in February 2012. For Sovereign¿ Compact systems manufactured at the AMO Manufacturing USA, LLC facility in Milpitas, CA the serial number format is YYYY80XXX where Y is for the year manufactured and "80" designates manufacturing at Milpitas. The Milpitas facility started manufacturing the v 5.3 systems in February 2013. The serial number is contained in the Product Identification Label (part number Z352548) that is attached to the rear of the console for systems manufactured.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1670-2013
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Abbott Medical Optics, Inc.? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Abbott Medical Optics, Inc.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Abbott Medical Optics, Inc. →

Source: openFDA record for Z-1670-2013. Report a data issue: hello@regdatalab.com.