Z-1670-2013 Class II Terminated
Recalled by Abbott Medical Optics, Inc.
— Milpitas, CA
FDA device recall Z-1670-2013 was initiated by Abbott Medical Optics, Inc. on May 23, 2013 and is designated Class II. Reason for recall: Abbott Medical Optics Inc. (AMO) initiated this Urgent Notice of Medical Device Correction and Advisory Notice to all customers who use the AMO Sovereign¿ Compact Phacoemulsification System, Version… The recall status is terminated (terminated Oct 6, 2014). Affected quantity: 534 systems.
Recall details
- Product Type
- Devices
- Report Date
- Jul 17, 2013
- Initiation Date
- May 23, 2013
- Termination Date
- Oct 6, 2014
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 534 systems
Product description
Sovereign¿ Compact Phacoemulsification System, v 5.3, Part Number: CMP680300. The system is used for cataract surgery.
Reason for recall
Abbott Medical Optics Inc. (AMO) initiated this Urgent Notice of Medical Device Correction and Advisory Notice to all customers who use the AMO Sovereign¿ Compact Phacoemulsification System, Version 5.3, because it has the potential to lose the vacuum tone mode during surgery.
Distribution pattern
Worldwide Distribution, including Nationwide (US) and countries in Latin America, EMEA, and APAC.
Code information
Product Name: Sovereign¿ Compact Console Universal, Part Number: CMP680300. The Sovereign¿ Compact console is identified with a unique, sequentially assigned 3 digit number identified below as XXX. For Sovereign¿ Compact systems manufactured at the Third Party Manufacturer, Sanmina-SCI Corporation, the serial number format is YYYY35XXX where Y is for the year manufactured and "35" designates manufacturing at Sanmina-SCI Corporation. The Sanmina-SCI Corporation started manufacturing the Sovereign¿ Compact v 5.3 systems in February 2012. For Sovereign¿ Compact systems manufactured at the AMO Manufacturing USA, LLC facility in Milpitas, CA the serial number format is YYYY80XXX where Y is for the year manufactured and "80" designates manufacturing at Milpitas. The Milpitas facility started manufacturing the v 5.3 systems in February 2013. The serial number is contained in the Product Identification Label (part number Z352548) that is attached to the rear of the console for systems manufactured.
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-1670-2013
- openFDA file date
- Nov 25, 2026
- Loaded by RegDataLab
- Nov 26, 2026
These dates say when openFDA published the data file and when RegDataLab last loaded it.
Need the full FDA record for Abbott Medical Optics, Inc.? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.
Question to ask in Claude
Use RegDataLab to give me an FDA risk summary for Abbott Medical Optics, Inc.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.
Ask Claude about Abbott Medical Optics, Inc. →
Use it in ChatGPT
Get a free API key
How the connector works