Z-0068-2018 Class II

Recalled by Limacorporate S.p.A — San Daniele del Friuli, N/A

Recall Details

Product Type
Devices
Report Date
November 15, 2017
Initiation Date
September 8, 2017
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
20 units in (US)

Product Description

Delta Cup Beater  Positioner - Aligner, model # 9057.20.555 Used during implantation of the Delta Acetabular System

Reason for Recall

An internal analysis on complaints received indicated that the welding process may have been performed sub optimally in the 12 lot # involved , leading to reduced mechanical strength of the device and possible breakage during impaction when the instrument is subjected to repeated multi-axial stresses.

Distribution Pattern

Worldwide Distribution - US Distribution and to the countries of : Australia, Austria, Belgium, Brazil, Czech Republic, Germany, Italy, New Zealand, Poland, Russia, Slovakia, Slovenia, Spain, South Africa, Switzerland, Turkey, UK and Germany

Code Information

Lot # 15AG0HJ, 15AG0HQ, 15AG0UY, 15AG00P, 15AG040, 15AG08Y, 14AG243, 15AG00Q, 15AG0UZ, 15AG0V0, 15AG0HN, 16AG02J