Z-0068-2018 Class II Terminated

Recalled by Limacorporate S.P.A — San Daniele del Friuli, N/A

FDA device recall Z-0068-2018 was initiated by Limacorporate S.P.A on Sep 8, 2017 and is designated Class II. Reason for recall: An internal analysis on complaints received indicated that the welding process may have been performed sub optimally in the 12 lot # involved , leading to reduced mechanical strength of the device and… The recall status is terminated (terminated Nov 19, 2026). Affected quantity: 20 units in (US).

Recall details

Product Type
Devices
Report Date
Nov 15, 2017
Initiation Date
Sep 8, 2017
Termination Date
Nov 19, 2026
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
20 units in (US)

Product description

Delta Cup Beater  Positioner - Aligner, model # 9057.20.555 Used during implantation of the Delta Acetabular System

Reason for recall

An internal analysis on complaints received indicated that the welding process may have been performed sub optimally in the 12 lot # involved , leading to reduced mechanical strength of the device and possible breakage during impaction when the instrument is subjected to repeated multi-axial stresses.

Distribution pattern

Worldwide Distribution - US Distribution and to the countries of : Australia, Austria, Belgium, Brazil, Czech Republic, Germany, Italy, New Zealand, Poland, Russia, Slovakia, Slovenia, Spain, South Africa, Switzerland, Turkey, UK and Germany

Code information

Lot # 15AG0HJ, 15AG0HQ, 15AG0UY, 15AG00P, 15AG040, 15AG08Y, 14AG243, 15AG00Q, 15AG0UZ, 15AG0V0, 15AG0HN, 16AG02J

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0068-2018
openFDA file date
Nov 24, 2026
Loaded by RegDataLab
Nov 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Limacorporate S.P.A? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-0068-2018. Report a data issue: hello@regdatalab.com.