Z-0086-2021 Class II Ongoing

Recalled by Roche Diagnostics Operations, Inc. — Indianapolis, IN

FDA device recall Z-0086-2021 was initiated by Roche Diagnostics Operations, Inc. on Aug 31, 2020 and is designated Class II. Reason for recall: Potential for Fluid leaks within the instruments, creating an electrical short inside the power socket, which generates heat, smoke, and eventually burning and melting of the electrical cord connector… The recall status is ongoing. Affected quantity: 257 (US); 208 (ex-US).

Recall details

Product Type
Devices
Report Date
Oct 14, 2020
Initiation Date
Aug 31, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
257 (US); 208 (ex-US)

Product description

DISCOVERY ULTRA Advanced Staining System (RUO-Research Use Only) Material Number: 05987750001

Reason for recall

Potential for Fluid leaks within the instruments, creating an electrical short inside the power socket, which generates heat, smoke, and eventually burning and melting of the electrical cord connector and socket

Distribution pattern

Nationwide Foreign: Canada, Austria, Belgium, Bulgaria, Croatia, Cyprus, Czech Republic, Denmark, Estonia, Finland, France, Germany, Greece, Hungary, Ireland, Italy, Latvia, Lithuania, Luxembourg, Netherlands, Norway, Poland, Portugal, Romania, Slovakia, Slovenia, Spain, Sweden, Switzerland, Turkey, UK, Algeria, Argentina, Aruba, Australia, Bahrain, Bosnia-Herz., Brazil, Chile, China, Colombia, Costa Rica, Cuba, Dominican Republic, Ecuador, Egypt, El Salvador, Hong Kong, India, Israel, Japan, Jordan, Kazakhstan, Kuwait, Malaysia, Martinque, Mexico, Monaco, Morocco, Namibia, New Caledonia, New Zealand, Oman, Pakistan, Panama, Peru, Philippines, Russia, Saudi Arabia, Singapore, S. Africa, S. Korea, Taiwan, Thailand, Ukraine, UAE, Vietnam

Code information

All units

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0086-2021
openFDA file date
Nov 24, 2026
Loaded by RegDataLab
Nov 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Roche Diagnostics Operations, Inc.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-0086-2021. Report a data issue: hello@regdatalab.com.