Z-0103-2021 Class III Terminated

Recalled by Roche Diagnostics Operations, Inc. — Indianapolis, IN

FDA device recall Z-0103-2021 was initiated by Roche Diagnostics Operations, Inc. on Aug 27, 2020 and is designated Class III. Reason for recall: Roche has confirmed customer complaints of low quality (QC) recovery and invalid calibrations for the identified products. A discoloration (yellow color) was observed and is the result of cross… The recall status is terminated (terminated May 22, 2024). Affected quantity: 180 units.

Recall details

Product Type
Devices
Report Date
Oct 21, 2020
Initiation Date
Aug 27, 2020
Termination Date
May 22, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
180 units

Product description

Tina-quant Complement C4 ver.2, Catalog 05991994190

Reason for recall

Roche has confirmed customer complaints of low quality (QC) recovery and invalid calibrations for the identified products. A discoloration (yellow color) was observed and is the result of cross contamination between the R1 and R3 reagents.

Distribution pattern

Domestic distribution nationwide.

Code information

lot 36870301