Z-0103-2021 Class III Terminated

Recalled by Roche Diagnostics Operations, Inc. — Indianapolis, IN

Recall Details

Product Type
Devices
Report Date
October 21, 2020
Initiation Date
August 27, 2020
Termination Date
May 22, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
180 units

Product Description

Tina-quant Complement C4 ver.2, Catalog 05991994190

Reason for Recall

Roche has confirmed customer complaints of low quality (QC) recovery and invalid calibrations for the identified products. A discoloration (yellow color) was observed and is the result of cross contamination between the R1 and R3 reagents.

Distribution Pattern

Domestic distribution nationwide.

Code Information

lot 36870301