Z-0105-2015 Class II Terminated
FDA device recall Z-0105-2015 was initiated by Philips Medical Systems, Inc. on Oct 3, 2014 and is designated Class II. Reason for recall: Philips Intellivue and Avalon Fetal Monitors in time-synchronized automatic/sequence mode, the NBP automatic measurement series is stopped The recall status is terminated (terminated Jan 31, 2017). Affected quantity: 118 units.
Recall details
- Product Type
- Devices
- Report Date
- Oct 29, 2014
- Initiation Date
- Oct 3, 2014
- Termination Date
- Jan 31, 2017
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 118 units
Product description
Reason for recall
Distribution pattern
Worldwide Distribution-USA (nationwide) and the countries of AUSTRALIA AUSTRIA BAHRAIN BELGIUM CHINA COLOMBIA CZECH REPUBLIC DENMARK FINLAND FRANCE GERMANY HONG KONG INDIA IRELAND ISRAEL ITALY JAPAN KAZAKHSTAN NETHERLANDS NEW ZEALAND NICARAGUA NIGERIA NORWAY OMAN POLAND ROMANIA RUSSIA SINGAPORE SLOVENIA SOUTH AFRICA SOUTH KOREA SPAIN SWEDEN SWITZERLAND TAIWAN THAILAND TURKEY UKRAINE and UNITED KINGDOM.
Code information
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-0105-2015
- openFDA file date
- Nov 27, 2026
- Loaded by RegDataLab
- Nov 28, 2026
Need the full FDA record for Philips Medical Systems, Inc.? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.
Question to ask in Claude
Use RegDataLab to give me an FDA risk summary for Philips Medical Systems, Inc.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.