Z-0110-2013 Class III Terminated

Recalled by Beckman Coulter Inc. — Brea, CA

FDA device recall Z-0110-2013 was initiated by Beckman Coulter Inc. on Dec 15, 2011 and is designated Class III. Reason for recall: The recall was initiated because Beckman Coulter has confirmed a mislabeling error, some Synchron Systems Multi Calibrator Diskettes (REF 442600, lot M002642G) may actually be Thyroxine (T4) Reagent… The recall status is terminated (terminated Dec 28, 2012). Affected quantity: 5746 units total (2729 units in US).

Recall details

Product Type
Devices
Report Date
Oct 31, 2012
Initiation Date
Dec 15, 2011
Termination Date
Dec 28, 2012
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
5746 units total (2729 units in US)

Product description

Synchron Systems Synchron Calibrator Multi Calibrator Diskettes, Part Number 442600. The SYNCHRON MULTI-CALIBRATOR, used in conjunction with SYNCHRON reagents, is intended for use on SYNCHRON Systems for the calibration of Albumin, Blood Urea Nitrogen (Urea), Calcium, Cholesterol, Glucose, Inorganic Phosphorus, Lactate,Magnesium, Total Protein, Triglycerides, and Uric Acid.

Reason for recall

The recall was initiated because Beckman Coulter has confirmed a mislabeling error, some Synchron Systems Multi Calibrator Diskettes (REF 442600, lot M002642G) may actually be Thyroxine (T4) Reagent Calibrator Diskettes.

Distribution pattern

Worldwide distribution: USA (Nationwide) and Internationally to: Afghanistan Angola Australia Bangladesh Belgium Bhutan Brazil Brunei Darussalam Bulgaria Canada China Colombia Costa Rica Cote d'Ivoire Czech Republic Denmark Ecuador Egypt El Salvador Eritrea Ethiopia Finland France French Polynesia Germany Ghana Greece Guatemala Hong Kong India Indonesia Ireland Israel Italy Japan Jordan Kenya Kuwait Lebanon Lesotho Libyan Arab Jamahiriya Lithuania Macao Malawi Malaysia Mexico Morocco Namibia Netherlands New Zealand Nigeria Oman Pakistan Panama Paraguay Philippines Poland Puerto Rico Qatar Reunion Romania Russian Federation Rwanda Saudi Arabia Serbia Singapore Slovakia South Africa Spain Swaziland Sweden Switzerland Taiwan Thailand Tunisia Turkey Uganda United Arab Emirates United Kingdom Viet Nam Zambia Zimbabwe

Code information

Lot M002642

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0110-2013
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-0110-2013. Report a data issue: hello@regdatalab.com.