Z-0121-2022 Class II Ongoing

Recalled by Fort Defiance Industries, LLC — Loudon, TN

FDA device recall Z-0121-2022 was initiated by Fort Defiance Industries, LLC on August 19, 2021 and is designated Class II. Reason for recall: The rear heater plate and cover assembly that contains the electrical wiring for the heaters does not have sufficient bonding to ground. The recall status is ongoing. Affected quantity: 380 devices.

Recall Details

Product Type
Devices
Report Date
October 27, 2021
Initiation Date
August 19, 2021
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
380 devices

Product Description

P2131 Automated Field Steam Sterilizer NSN: 6530-01-641-4641

Reason for Recall

The rear heater plate and cover assembly that contains the electrical wiring for the heaters does not have sufficient bonding to ground.

Distribution Pattern

Distributed nationwide to PA, MD, TX, NJ, OH, TN, UT, VA and internationally to Canada.

Code Information

Model number: P2131; NSN: 6530-01-641-4641; Serial Numbers AFS-0001 through AFS-0386. Note: AFS-0297 was not manufactured.