Z-2983-2026 Class II Ongoing
FDA device recall Z-2983-2026 was initiated by Fort Defiance Industries, LLC on July 8, 2026 and is designated Class II. Reason for recall: The Automated Field Steam Sterilizer's water recovery system 120V Amphenol connector may experience an intermittent loss of ground continuity, which could result in an unreliable fault current return … The recall status is ongoing.
Recall Details
- Product Type
- Devices
- Report Date
- August 26, 2026
- Initiation Date
- July 8, 2026
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
Product Description
Fort Defiance Industries (FDI), Automated Field Steam Sterilizer REF: P2131
Reason for Recall
The Automated Field Steam Sterilizer's water recovery system 120V Amphenol connector may experience an intermittent loss of ground continuity, which could result in an unreliable fault current return path.
Distribution Pattern
Worldwide - US Nationwide distribution in the states of MD, VA, NJ, TN and the countries of Canada, Lithuania, Spain.
Code Information
Lot Code: All serial numbers/ GTIN: (01) 00862992000327