Z-2983-2026 Class II Ongoing
FDA device recall Z-2983-2026 was initiated by Fort Defiance Industries, LLC on Jul 8, 2026 and is designated Class II. Reason for recall: The Automated Field Steam Sterilizer's water recovery system 120V Amphenol connector may experience an intermittent loss of ground continuity, which could result in an unreliable fault current return… The recall status is ongoing.
Recall details
- Product Type
- Devices
- Report Date
- Aug 26, 2026
- Initiation Date
- Jul 8, 2026
- Voluntary/Mandated
- Voluntary: Firm initiated
Product description
Reason for recall
Distribution pattern
Worldwide - US Nationwide distribution in the states of MD, VA, NJ, TN and the countries of Canada, Lithuania, Spain.
Code information
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-2983-2026
- openFDA file date
- Nov 19, 2026
- Loaded by RegDataLab
- Nov 20, 2026
Need the full FDA record for Fort Defiance Industries, LLC? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.