Z-2983-2026 Class II Ongoing

Recalled by Fort Defiance Industries, LLC — Loudon, TN

FDA device recall Z-2983-2026 was initiated by Fort Defiance Industries, LLC on July 8, 2026 and is designated Class II. Reason for recall: The Automated Field Steam Sterilizer's water recovery system 120V Amphenol connector may experience an intermittent loss of ground continuity, which could result in an unreliable fault current return … The recall status is ongoing.

Recall Details

Product Type
Devices
Report Date
August 26, 2026
Initiation Date
July 8, 2026
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity

Product Description

Fort Defiance Industries (FDI), Automated Field Steam Sterilizer REF: P2131

Reason for Recall

The Automated Field Steam Sterilizer's water recovery system 120V Amphenol connector may experience an intermittent loss of ground continuity, which could result in an unreliable fault current return path.

Distribution Pattern

Worldwide - US Nationwide distribution in the states of MD, VA, NJ, TN and the countries of Canada, Lithuania, Spain.

Code Information

Lot Code: All serial numbers/ GTIN: (01) 00862992000327