Z-2983-2026 Class II Ongoing

Recalled by Fort Defiance Industries, LLC — Loudon, TN

FDA device recall Z-2983-2026 was initiated by Fort Defiance Industries, LLC on Jul 8, 2026 and is designated Class II. Reason for recall: The Automated Field Steam Sterilizer's water recovery system 120V Amphenol connector may experience an intermittent loss of ground continuity, which could result in an unreliable fault current return… The recall status is ongoing.

Recall details

Product Type
Devices
Report Date
Aug 26, 2026
Initiation Date
Jul 8, 2026
Voluntary/Mandated
Voluntary: Firm initiated

Product description

Fort Defiance Industries (FDI), Automated Field Steam Sterilizer REF: P2131

Reason for recall

The Automated Field Steam Sterilizer's water recovery system 120V Amphenol connector may experience an intermittent loss of ground continuity, which could result in an unreliable fault current return path.

Distribution pattern

Worldwide - US Nationwide distribution in the states of MD, VA, NJ, TN and the countries of Canada, Lithuania, Spain.

Code information

Lot Code: All serial numbers/ GTIN: (01) 00862992000327

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2983-2026
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Fort Defiance Industries, LLC? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-2983-2026. Report a data issue: hello@regdatalab.com.