Z-2411-2018 Class II Terminated

Recalled by Fort Defiance Industries, LLC — Loudon, TN

Recall Details

Product Type
Devices
Report Date
July 18, 2018
Initiation Date
May 24, 2018
Termination Date
June 27, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
199 units

Product Description

FDI FORT DEFIANCE INDUSTRIES P2131 AUTOMATED FIELD STEAM STERILIZER, NSN: 6530-01-641-4641 Product Usage: The Fort Defiance Industries Automated Field Steam Sterilizer Model P2131 is designed for sterilization of porous and non-porous, heat- and moisture-stable materials (e.g., surgical instruments and textiles) used in healthcare facilities. The P2131 sterilizer is a transportable device designed for use in a variety of austere environments.

Reason for Recall

During long-term storage (i.e. military depot), the interaction of stagnant water and dissimilar metals in the plumbing of the P2131 device produces deposits in the water system that can potentially interfere with the rear sight gauge and / or the water liquid level switch.

Distribution Pattern

US Nationwide Distribution in the states of PA. VA. TX

Code Information

Serial Numbers: AFS-0001 through AFS-0199