Z-0143-2023 Class I Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- November 9, 2022
- Initiation Date
- September 30, 2022
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 1137 units
Product Description
Hillrom Hospital Bed Accessory, Watch Care Incontinence Management System for Progressa Bed, Product Code/Part Number P00697903.
Reason for Recall
Radio frequency emission from a functioning WatchCare device may potentially impact other devices (including, but not limited to telemetry devices, bladder scanner, fetal monitor/doppler, infusion pumps, and insulin pumps).
Distribution Pattern
US
Code Information
UDI/DI 00887761998889, all serial numbers