Z-0172-2018 Class II Terminated
Recalled by Brainlab AG
— Munich, N/A
FDA device recall Z-0172-2018 was initiated by Brainlab AG on Nov 3, 2017 and is designated Class II. Reason for recall: Brainlab has internally detected that under specific conditions ExacTrac Vero may not correctly account for this ring angle correction during the calculation of the corresponding couch shift. This… The recall status is terminated (terminated Dec 14, 2020). Affected quantity: 28 units.
Recall details
- Product Type
- Devices
- Report Date
- Dec 6, 2017
- Initiation Date
- Nov 3, 2017
- Termination Date
- Dec 14, 2020
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 28 units
Product description
BRAINLAB EXACTRAC VERO, Model/Catalog Numbers: 46228 EXACTRAC VERO 3.5 46238 EXACTRAC VERO 3.5 46216 EXACTRAC VERO SW UPDATE 3.5.2 TO 3.5.3 46218 EXACTRAC VERO SW UPDATE 3.5.3 TO 3.5.4 The ExacTrac Vero system is intended to be used in conjunction with the MHI-TM2000 Radiation Therapy Linear Accelerator System manufactured by Mitsubishi Heavy Industries, Ltd. ExacTrac Vero uses the images received from the MHI-TM2000 linear accelerator for analyzing the current patient position and calculating - when applicable - a necessary correction shift. The correction shift is then exported to the MHI-TM2000 linear accelerator. The ExacTrac Vero system uses stereoscopic X-ray or Cone Beam CT registration and optical tracking of infrared reflective markers in order to localize and correct the patient position before and during treatment. Optionally ExacTrac Vero provides position data for the pan/tilt motion of the MHI-TM2000 gantry head to the MHITM2000 controller for continuous alignment of the beam orientation with a moving target. The position data is based on target detection via X-ray imaging and IR tracking of external surrogate markers.
Reason for recall
Brainlab has internally detected that under specific conditions ExacTrac Vero may not correctly account for this ring angle correction during the calculation of the corresponding couch shift. This may result in the treatment couch not moving to the exact planned position resulting in a deviation between the planned and the treatment target position. This deviation may not be clearly visible to the user if no verification image of the patient position is acquired after this couch shift has been performed.
Distribution pattern
Worldwide Distribution - US Distribution and to the countries of : Belgium, Canada, France, Germany, Italy, Japan and S. Korea
Code information
Potentially affected are ExacTrac Vero versions 3.x. Currently only ExacTrac Vero SW versions 3.5(.x) are installed for use on existing ExacTrac Vero Systems (only ExacTrac Vero v.3.5 exist in the field). UDI: 04056481132446
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-0172-2018
- openFDA file date
- Nov 24, 2026
- Loaded by RegDataLab
- Nov 25, 2026
These dates say when openFDA published the data file and when RegDataLab last loaded it.
Need the full FDA record for Brainlab AG? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.
Get a free API key
How the connector works