Z-0181-2015 Class II Terminated
FDA device recall Z-0181-2015 was initiated by Philips Medical Systems, Inc. on Jul 8, 2014 and is designated Class II. Reason for recall: Faulty Automatic Motion Controller (AMC), a problem in the Power On Self Test (POST) error handling was detected, can result in a hazardous movement of the C-arc. system. The recall status is terminated (terminated Jan 18, 2017). Affected quantity: 22 systems.
Recall details
- Product Type
- Devices
- Report Date
- Nov 19, 2014
- Initiation Date
- Jul 8, 2014
- Termination Date
- Jan 18, 2017
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 22 systems
Product description
Reason for recall
Distribution pattern
Distributed in the states of IL, and TN, and the countries of Argentina , Australia, Belgium, Denmark, Egypt, France, Germany, Hong Kong, Italy, Japan, Korea, Republic of Netherlands, Norway, Saudi Arabia, Switzerland, United Kingdom, and Viet Nam.
Code information
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-0181-2015
- openFDA file date
- Nov 27, 2026
- Loaded by RegDataLab
- Nov 28, 2026
Need the full FDA record for Philips Medical Systems, Inc.? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.
Question to ask in Claude
Use RegDataLab to give me an FDA risk summary for Philips Medical Systems, Inc.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.