Z-0252-2013 Class II Terminated
FDA device recall Z-0252-2013 was initiated by Diagnostica Stago, Inc. on Jun 15, 2012 and is designated Class II. Reason for recall: Following a single customer complaint that incorrect INR values were occasionally printed for PT tests, even though the primary units (seconds) were correct, the firm investigated and confirmed the… The recall status is terminated (terminated Jul 19, 2013). Affected quantity: 232 units US; 5 units Canada.
Recall details
- Product Type
- Devices
- Report Date
- Nov 14, 2012
- Initiation Date
- Jun 15, 2012
- Termination Date
- Jul 19, 2013
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 232 units US; 5 units Canada
Product description
STA Satellite Automated Multi-Parameter Analyzer This system is a fully automatic clinical instrument indicated and intended for the performance of tests on human plasma, the results of which aid in the diagnosis of coagulation abnormalities or in monitoring anticoagulant therapy.
Reason for recall
Following a single customer complaint that incorrect INR values were occasionally printed for PT tests, even though the primary units (seconds) were correct, the firm investigated and confirmed the potential for printing incorrect secondary units when using an external printer and the customer printout mode on the STA Satellite¿.
Distribution pattern
Worldwide Distribution - USA (nationwide) and Canada
Code information
K082248 D089976 Ref 58104 units