Z-0297-2026 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- October 29, 2025
- Initiation Date
- September 12, 2025
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 1386 units
Product Description
Millenium MG/MC; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging
Reason for Recall
GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
Distribution Pattern
Worldwide
Code Information
all serial numbers in distribution