Z-0300-2026 Class II Ongoing

Recalled by GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING — Haifa

FDA device recall Z-0300-2026 was initiated by GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING on Sep 12, 2025 and is designated Class II. Reason for recall: GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service could have been transported or relocated without adequate detector support. This could… The recall status is ongoing. Affected quantity: 54 units.

Recall details

Product Type
Devices
Report Date
Oct 29, 2025
Initiation Date
Sep 12, 2025
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
54 units

Product description

Millenium MPS; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging

Reason for recall

GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.

Distribution pattern

Worldwide

Code information

all serial numbers in distribution

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0300-2026
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-0300-2026. Report a data issue: hello@regdatalab.com.