Z-0334-2024 Class II Ongoing

Recalled by Aizu Olympus Co., Ltd. — Aizuwakamatsu, N/A

Recall Details

Product Type
Devices
Report Date
November 29, 2023
Initiation Date
September 25, 2023
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
3627 units

Product Description

Gastrointestinal videoscope-indicated for use within the upper digestive tract (including the esophagus, stomach, and duodenum). Model Numbers: GIF-1TH190, GIF-1TQ160, GIF-2T160, GIF-2TH180, GIF-H170, GIF-H180, GIF-H180J, GIF-H190, GIF-HQ190, GIF-LV1, GIF-N180, GIF-Q150, GIF-Q160Z, GIF-Q180, GIF-XP180N, GIF-XP190N, GIF-XTQ160. 510(k) Numbers: K112680, K954451, K051645, K100584, K111756, K011151, K123317.

Reason for Recall

Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection

Distribution Pattern

Worldwide - US Nationwide distribution.

Code Information

Model Numbers: GIF-1TH190 (04953170343360), GIF-1TQ160 (Not recorded), GIF-2T160 (04953170307492), GIF-2TH180 (04953170339745), GIF-H170 (04953170334115), GIF-H180 (04953170339769), GIF-H180J (04953170307539), GIF-H190 (04953170305290), GIF-HQ190 (04953170305276), GIF-LV1 (04953170312373), GIF-N180 (04953170307508), GIF-Q150 (04953170307485), GIF-Q160Z (04953170066795), GIF-Q180 (04953170307522), GIF-XP180N (04953170307515), GIF-XP190N (04953170305313), GIF-XTQ160 (04953170339837).