Z-0340-2024 Class II Ongoing

Recalled by Aizu Olympus Co., Ltd. — Aizuwakamatsu, N/A

Recall Details

Product Type
Devices
Report Date
November 29, 2023
Initiation Date
September 25, 2023
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
33 units

Product Description

Mobile airway scope-diagnosis and observation to access airway anatomy, endotracheal/endobronchial intubation and management. Model Numbers: MAF-DM2, MAF-GM, MAF-GM2, MAF-TM, MAF-TM2.

Reason for Recall

Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection

Distribution Pattern

Worldwide - US Nationwide distribution.

Code Information

Model Numbers: MAF-DM2 (04953170407260), MAF-GM (04953170340369), MAF-GM2 (04953170407208), MAF-TM (04953170288630), MAF-TM2 (04953170407147).