Z-0345-2024 Class II Ongoing

Recalled by Aizu Olympus Co., Ltd. — Aizuwakamatsu, N/A

Recall Details

Product Type
Devices
Report Date
November 29, 2023
Initiation Date
September 25, 2023
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
5 units

Product Description

Ultrasonic gastrovideoscope-Real time ultrasound imaging,ultrasound guided needle aspiration and other endoscopic procedures within the upper gastrointestinal tract and surrounding organs Model Number: TGF-UC180J.

Reason for Recall

Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection

Distribution Pattern

Worldwide - US Nationwide distribution.

Code Information

Model Number (UDI-DI): TGF-UC180J (04953170328480).