Z-0347-2024 Class II Ongoing

Recalled by Aizu Olympus Co., Ltd. — Aizuwakamatsu, N/A

FDA device recall Z-0347-2024 was initiated by Aizu Olympus Co., Ltd. on Sep 25, 2023 and is designated Class II. Reason for recall: Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the… The recall status is ongoing. Affected quantity: 1577 units.

Recall details

Product Type
Devices
Report Date
Nov 29, 2023
Initiation Date
Sep 25, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1577 units

Product description

Ureteralscope- for endoscopic diagnosis and treatment within the ureter, kidney and biliary tract (common bile duct and hepatic duct). Model Numbers: URF-P6, URF-P6R.

Reason for recall

Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection

Distribution pattern

Worldwide - US Nationwide distribution.

Code information

Model Number (UDI-DI): URF-P6 (04953170340802), URF-P6R (04953170340833).

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)