Z-0389-2014 Class II Terminated

Recalled by Mindray DS USA, Inc. Dba Mindray North America — Mahwah, NJ

FDA device recall Z-0389-2014 was initiated by Mindray DS USA, Inc. Dba Mindray North America on Oct 7, 2013 and is designated Class II. Reason for recall: Mindray has identified an issue with V Series Monitor where the V Dock power pins connecting the V Series Monitor to the V Dock may lose spring tension, potentially causing the screen to temporarily… The recall status is terminated (terminated Apr 27, 2015). Affected quantity: 1839 units.

Recall details

Product Type
Devices
Report Date
Dec 4, 2013
Initiation Date
Oct 7, 2013
Termination Date
Apr 27, 2015
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1839 units

Product description

Multi Parameter Patient Monitor (with Arrhythmia Detection and Alarms) Product Usage: The V Series Monitor is intended for intra hospital use under the direct supervision of a licensed healthcare practitioner. The indications for use for the V Series include the monitoring of the following human physiological parameters: 1) ECG waveform derived from 3, 5, 6 and 12 lead measurements 2) Heart Rate 3) Pulse Oximtery 4) ST Segment 5) Arrhythmia Detection 6) Non Invasive Blood Pressure (IBP) 7) Cardiac Output (CO) 8) Respiratory Gases 9) Respiration Rate 10) Temperature The V series monitor has the capability of performing IV Drug and Hemodynamic Calculations and interfacing with network devices.

Reason for recall

Mindray has identified an issue with V Series Monitor where the V Dock power pins connecting the V Series Monitor to the V Dock may lose spring tension, potentially causing the screen to temporarily go blank, lose communication with the central station monitor (when one is in use) or shut down. No patient related events have been reported to Mindray concerning the issue.

Distribution pattern

Worldwide Distribution - USA Nationwide

Code information

V Series Montor p/n 0998-00-1801-01 and the associated V Dock , p/n 0998-UC-180-01

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0389-2014
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-0389-2014. Report a data issue: hello@regdatalab.com.