Z-0389-2014 Class II Terminated
FDA device recall Z-0389-2014 was initiated by Mindray DS USA, Inc. Dba Mindray North America on Oct 7, 2013 and is designated Class II. Reason for recall: Mindray has identified an issue with V Series Monitor where the V Dock power pins connecting the V Series Monitor to the V Dock may lose spring tension, potentially causing the screen to temporarily… The recall status is terminated (terminated Apr 27, 2015). Affected quantity: 1839 units.
Recall details
- Product Type
- Devices
- Report Date
- Dec 4, 2013
- Initiation Date
- Oct 7, 2013
- Termination Date
- Apr 27, 2015
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 1839 units
Product description
Multi Parameter Patient Monitor (with Arrhythmia Detection and Alarms) Product Usage: The V Series Monitor is intended for intra hospital use under the direct supervision of a licensed healthcare practitioner. The indications for use for the V Series include the monitoring of the following human physiological parameters: 1) ECG waveform derived from 3, 5, 6 and 12 lead measurements 2) Heart Rate 3) Pulse Oximtery 4) ST Segment 5) Arrhythmia Detection 6) Non Invasive Blood Pressure (IBP) 7) Cardiac Output (CO) 8) Respiratory Gases 9) Respiration Rate 10) Temperature The V series monitor has the capability of performing IV Drug and Hemodynamic Calculations and interfacing with network devices.
Reason for recall
Mindray has identified an issue with V Series Monitor where the V Dock power pins connecting the V Series Monitor to the V Dock may lose spring tension, potentially causing the screen to temporarily go blank, lose communication with the central station monitor (when one is in use) or shut down. No patient related events have been reported to Mindray concerning the issue.
Distribution pattern
Worldwide Distribution - USA Nationwide
Code information
V Series Montor p/n 0998-00-1801-01 and the associated V Dock , p/n 0998-UC-180-01