Z-0422-2023 Class II Ongoing

Recalled by Getinge Usa Sales Inc — Wayne, NJ

FDA device recall Z-0422-2023 was initiated by Getinge Usa Sales Inc on Nov 7, 2022 and is designated Class II. Reason for recall: The power backup battery, under certain conditions, may exhibit reduced battery run-time while in use, and there is a potential that the battery may fail to provide the expected run time in the event… The recall status is ongoing. Affected quantity: 65 US.

Recall details

Product Type
Devices
Report Date
Dec 14, 2022
Initiation Date
Nov 7, 2022
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
65 US

Product description

Flow-c Anesthesia System (Product Code 6887700) Flow-e Anesthesia System (Product Code 6887900)

Reason for recall

The power backup battery, under certain conditions, may exhibit reduced battery run-time while in use, and there is a potential that the battery may fail to provide the expected run time in the event of a mains power failure. This could require the clinician to use manual emergency ventilation of another ventilator is not available and the backup battery is exhausted.

Distribution pattern

US Nationwide distribution in the states of CO, FL, IN, MA, MO, and VA.

Code information

Flow-c UDI-DI: 7325710010457 Lot/Serial Numbers 4003 4777 4783 5024 5025 5026 5027 5028 5043 5044 5045 5046 5047 5048 5049 5050 5078 Flow-e UDI-DI: 7325710009765 Lot/Serial Numbers 50012 50013 50014 50015 50016 50017 50030 50031 50033 50034 50035 50036 50037 50038 50039 50040 50041 50042 50043 50046 50047 50048 50051 50052 50053 50054 50055 50056 50057 50058 50059 50060 50061 50062 50069 50070 50071 50072 50073 50074 50084 50085 50086 50087 50088 50089 50090 50091 50092 50093 50094 50096 50097 50108 50109 50110 50111 50132

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0422-2023
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-0422-2023. Report a data issue: hello@regdatalab.com.