Z-0445-2022 Class II Ongoing
FDA device recall Z-0445-2022 was initiated by Invacare Corporation on Nov 12, 2021 and is designated Class II. Reason for recall: Power Wheelchairs with a LiNX Gyro component running firmware version 6.1.2 can experience a more aggressive deceleration rate than the programmed rate due to the system following the incorrect… The recall status is ongoing. Affected quantity: 1477 units US; 1226 units OUS.
Recall details
- Product Type
- Devices
- Report Date
- Jan 12, 2022
- Initiation Date
- Nov 12, 2021
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 1477 units US; 1226 units OUS
Product description
Invacare TDX SP2 with LiNX Gyro Power Wheelchair (with LiNX Gyro Module Model Numbers DLX-GRY100-A and DLX-GRY101-A) Models: TDX SP2: TDXSP2, TDXSP2-CG, TDXSP2X-CG, TDXSP2-MCG, TDXSP2X-MCG, TDXSP2V, TDXSP2V-HD, TDXSP2HD, TDXSP2HD-CG, TDXSP2HD-MCG
Reason for recall
Power Wheelchairs with a LiNX Gyro component running firmware version 6.1.2 can experience a more aggressive deceleration rate than the programmed rate due to the system following the incorrect deceleration profile, resulting in potential injury
Distribution pattern
Worldwide distribution - US Nationwide and the countries of Canada, Australia, Thailand, New Zealand.
Code information
LiNX Gyro firmware version 6.1.2 UDI-DI Numbers: TDXSP2 00841447100911 TDXSP2-CG 00841447100942 TDXSP2-MCG 00841447100966 TDXSP2HD 00841447108641 TDXSP2HD-CG 00841447111207 TDXSP2HD-MCG 00841447111214 TDXSP2V 00841447101512 TDXSP2V-HD 00841447102052 TDXSP2X-CG 00841447101000 TDXSP2X-MCG 00841447101017
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-0445-2022
- openFDA file date
- Nov 10, 2026
- Loaded by RegDataLab
- Nov 11, 2026