Z-0446-2022 Class II Ongoing
FDA device recall Z-0446-2022 was initiated by Invacare Corporation on Nov 12, 2021 and is designated Class II. Reason for recall: Power Wheelchairs with a LiNX Gyro component running firmware version 6.1.2 can experience a more aggressive deceleration rate than the programmed rate due to the system following the incorrect… The recall status is ongoing. Affected quantity: 303 units US; 20 units OUS.
Recall details
- Product Type
- Devices
- Report Date
- Jan 12, 2022
- Initiation Date
- Nov 12, 2021
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 303 units US; 20 units OUS
Product description
Invacare AVIVA FX with LiNX Gyro Power Wheelchair (with LiNX Gyro Module Model Numbers DLX-GRY100-A and DLX-GRY101-A) Models: IFX-20R, IFX-20SP, IFX-20MP, IFX-20C
Reason for recall
Power Wheelchairs with a LiNX Gyro component running firmware version 6.1.2 can experience a more aggressive deceleration rate than the programmed rate due to the system following the incorrect deceleration profile, resulting in potential injury
Distribution pattern
Worldwide distribution - US Nationwide and the countries of Canada, Australia, Thailand, New Zealand.
Code information
LiNX Gyro firmware version 6.1.2 UDI-DI Numbers: IFX-20C 00841447111627 IFX-20MP 00841447111658 IFX-20R 00841447111634 IFX-20SP 00841447111641 IFX20WIDE 00841447111658
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-0446-2022