Z-0447-2022 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- January 12, 2022
- Initiation Date
- November 12, 2021
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 1 unit US
Product Description
Invacare AVIVA Storm RX with LiNX Gyro Power Wheelchair (with LiNX Gyro Module Model Numbers DLX-GRY100-A and DLX-GRY101-A) Models: SRX-20R, SRX-20SP, SRX-20MP
Reason for Recall
Power Wheelchairs with a LiNX Gyro component running firmware version 6.1.2 can experience a more aggressive deceleration rate than the programmed rate due to the system following the incorrect deceleration profile, resulting in potential injury
Distribution Pattern
Worldwide distribution - US Nationwide and the countries of Canada, Australia, Thailand, New Zealand.
Code Information
LiNX Gyro firmware version 6.1.2 UDI-DI Numbers: SRX-20MP 00841447114581 SRX-20R 00841447114604 SRX-20SP 00841447114598