Z-0447-2022 Class II Ongoing

Recalled by Invacare Corporation — Elyria, OH

Recall Details

Product Type
Devices
Report Date
January 12, 2022
Initiation Date
November 12, 2021
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1 unit US

Product Description

Invacare AVIVA Storm RX with LiNX Gyro Power Wheelchair (with LiNX Gyro Module Model Numbers DLX-GRY100-A and DLX-GRY101-A) Models: SRX-20R, SRX-20SP, SRX-20MP

Reason for Recall

Power Wheelchairs with a LiNX Gyro component running firmware version 6.1.2 can experience a more aggressive deceleration rate than the programmed rate due to the system following the incorrect deceleration profile, resulting in potential injury

Distribution Pattern

Worldwide distribution - US Nationwide and the countries of Canada, Australia, Thailand, New Zealand.

Code Information

LiNX Gyro firmware version 6.1.2 UDI-DI Numbers: SRX-20MP 00841447114581 SRX-20R 00841447114604 SRX-20SP 00841447114598