Z-0447-2022 Class II Ongoing

Recalled by Invacare Corporation — Elyria, OH

FDA device recall Z-0447-2022 was initiated by Invacare Corporation on Nov 12, 2021 and is designated Class II. Reason for recall: Power Wheelchairs with a LiNX Gyro component running firmware version 6.1.2 can experience a more aggressive deceleration rate than the programmed rate due to the system following the incorrect… The recall status is ongoing. Affected quantity: 1 unit US.

Recall details

Product Type
Devices
Report Date
Jan 12, 2022
Initiation Date
Nov 12, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1 unit US

Product description

Invacare AVIVA Storm RX with LiNX Gyro Power Wheelchair (with LiNX Gyro Module Model Numbers DLX-GRY100-A and DLX-GRY101-A) Models: SRX-20R, SRX-20SP, SRX-20MP

Reason for recall

Power Wheelchairs with a LiNX Gyro component running firmware version 6.1.2 can experience a more aggressive deceleration rate than the programmed rate due to the system following the incorrect deceleration profile, resulting in potential injury

Distribution pattern

Worldwide distribution - US Nationwide and the countries of Canada, Australia, Thailand, New Zealand.

Code information

LiNX Gyro firmware version 6.1.2 UDI-DI Numbers: SRX-20MP 00841447114581 SRX-20R 00841447114604 SRX-20SP 00841447114598

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0447-2022
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-0447-2022. Report a data issue: hello@regdatalab.com.