Z-0461-2021 Class II Terminated
FDA device recall Z-0461-2021 was initiated by Roche Diagnostics Operations, Inc. on October 9, 2020 and is designated Class II. Reason for recall: Potential Incorrect Validation of Results Due to an Erroneous QC Status When Using Status Expiration Control Rule The recall status is terminated (terminated June 9, 2023). Affected quantity: 40 licenses.
Recall Details
- Product Type
- Devices
- Report Date
- November 25, 2020
- Initiation Date
- October 9, 2020
- Termination Date
- June 9, 2023
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 40 licenses
Product Description
cobas infinity central lab/cobas infinity core license-The cobas infinity IT solutions application is intended to be used for the configuration and connectivity management of instruments and software systems Catalog Number: 07154003001
Reason for Recall
Potential Incorrect Validation of Results Due to an Erroneous QC Status When Using Status Expiration Control Rule
Distribution Pattern
Nationwide
Code Information
Software versions from cobas infinity version 3.00.00 onwards Serial Numbers: 82014 82229 82028 82227 82066 81967 81946 82121 82027 82194 82120 81528 82231 81958 82246 81983 82118 82165 81886 82082 82256 82257 81781 82331 82304 82087 82276 81465 81959 82235 82180 82181 82236 81945 82275 81662 82065 82253 81817 82000