Z-0586-2021 Class II Terminated

Recalled by Synthes (USA) Products LLC — West Chester, PA

Recall Details

Product Type
Devices
Report Date
December 23, 2020
Initiation Date
November 19, 2020
Termination Date
July 1, 2022
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
515 US; 150 OUS

Product Description

DePuy Synthes 17.5MM REAMER HEAD FOR RIA 2 STERILE-Orthopedic manual surgical instrument Part Code: 03.404.031S

Reason for Recall

RIA 2 reamer head breakage exceeds the expected occurrence rate resulting in adverse tissue reaction, soft tissue damage, and/or bone damage, when the recommended surgical techniques are not followed

Distribution Pattern

Nationwide Foreign:countries: United Arab Emirates, Austria, Australia, Belgium, Switzerland, China, Czech Republic, Germany, Spain, United Kingdom, Ireland, Israel, Italy, Singapore,

Code Information

All Lots UDI: 10886982274076