Z-0634-2022 Class II Ongoing

Recalled by Synthes (USA) Products LLC — West Chester, PA

FDA device recall Z-0634-2022 was initiated by Synthes (USA) Products LLC on Dec 22, 2021 and is designated Class II. Reason for recall: Top cap may loosen and detach if the Universal Chuck becomes jammed and the user attempts to manually free the device, internal ball bearings may fall out onto the surgical field and cause surgical… The recall status is ongoing. Affected quantity: 600 units.

Recall details

Product Type
Devices
Report Date
Feb 16, 2022
Initiation Date
Dec 22, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
600 units

Product description

Universal Chuck-The Universal Chuck is an Instrument Handle provided as an alternative instrument available for use in the TN-ADVANCED" Tibial Nailing System (TNA), TFN-ADVANCED" Proximal Femoral Nailing System (TFNA), RFN-ADVANCED" Retrograde Femoral Nailing System (RFNA), FRN-ADVANCED" Femoral Recon Nailing System (FRNA), and Flexible Monobloc Reamers. Instrument Handles are intended for holding instruments for manipulation during orthopedic surgery. Product No.: 03.043.001

Reason for recall

Top cap may loosen and detach if the Universal Chuck becomes jammed and the user attempts to manually free the device, internal ball bearings may fall out onto the surgical field and cause surgical delay or, and if not recovered, adverse tissue reaction and infection.

Distribution pattern

US Nationwide distribution.

Code information

Lot/Serial No.: 20013101, 20052701, 20228901, 20229301, 20253901, 20558602, 20698102, 20777201 21869701, 21983904, 21995301 GTIN: 10886982297372

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0634-2022
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Synthes (USA) Products LLC? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Synthes (USA) Products LLC: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Synthes (USA) Products LLC →

Source: openFDA record for Z-0634-2022. Report a data issue: hello@regdatalab.com.