Z-0669-2020 Class II Terminated

Recalled by Roche Diagnostics Operations, Inc. — Indianapolis, IN

Recall Details

Product Type
Devices
Report Date
December 18, 2019
Initiation Date
October 28, 2019
Termination Date
June 28, 2022
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
712 units

Product Description

Elecsys Anti-HAV IgM assay-for the in vitro qualitative detection of IgM antibodies to hepatitis A virus (anti-HAV IgM) in human serum and plasma Catalog number: 07026773190

Reason for Recall

Current software version of the Anti-HAV IgM APP-file (version 06.05-101) downloaded an erroneous Border Area setting, (1.0 1.00) instead of (0.90 1.10).

Distribution Pattern

US Nationwide distribution.

Code Information

Anti-HAV IgM APP-file (version 06.05-101)