Z-0669-2020 Class II Terminated

Recalled by Roche Diagnostics Operations, Inc. — Indianapolis, IN

FDA device recall Z-0669-2020 was initiated by Roche Diagnostics Operations, Inc. on Oct 28, 2019 and is designated Class II. Reason for recall: Current software version of the Anti-HAV IgM APP-file (version 06.05-101) downloaded an erroneous Border Area setting, (1.0 1.00) instead of (0.90 1.10). The recall status is terminated (terminated Jun 28, 2022). Affected quantity: 712 units.

Recall details

Product Type
Devices
Report Date
Dec 18, 2019
Initiation Date
Oct 28, 2019
Termination Date
Jun 28, 2022
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
712 units

Product description

Elecsys Anti-HAV IgM assay-for the in vitro qualitative detection of IgM antibodies to hepatitis A virus (anti-HAV IgM) in human serum and plasma Catalog number: 07026773190

Reason for recall

Current software version of the Anti-HAV IgM APP-file (version 06.05-101) downloaded an erroneous Border Area setting, (1.0 1.00) instead of (0.90 1.10).

Distribution pattern

US Nationwide distribution.

Code information

Anti-HAV IgM APP-file (version 06.05-101)

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0669-2020
openFDA file date
Nov 24, 2026
Loaded by RegDataLab
Nov 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Roche Diagnostics Operations, Inc.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-0669-2020. Report a data issue: hello@regdatalab.com.